Published:Journal of Chromatographic Science,
ISSN 0021-9665 Volume
46, Number 8, September 2008, pp. 717-721
High-Sensitivity Simultaneous Analysis of Ubiquinol-10
and Ubiquinone-10 in Human Plasma
Aya Hirayama1, Hiroshi Kubo2,
Masashi Mita1,
Osamu Shirota1, and Yorihiro Yamamoto3, 1Shiseido Research Center, 2-2-1 Hayabuchi, Tsuzuki-ku, Yokohama
224-8558, Japan; 2Department of Chemistry and Biotechnology,
Graduate School of Engineering, The University of Tokyo, Hongo,
Bunkyo-ku, Tokyo 113-8656, Japan; 3School of Bionics, Tokyo University
of Technology, 1404-1 Katakura, Hachioji 192-0982, Japan
A method to determine ubiquinol-10 and ubiquinone-10
in human serum was developed by using high-performance liquid
chromatography consisting of a semi-microcolumn switching system
and an electrochemical detector (ECD), which requires minimized
sample pre-treatments. A linear dynamic range was obtained from
1.0 to 5000 ng/mL, and recovery values of 89–105% were
observed in a low-concentration region of 10–50 ng/mL.
In a long operation test, a good precision was maintained during
5100 runs without any maintenance on ECD or columns. In addition,
retention behaviors of other ubiquinone homologues were examined.
Reproduction
of editorial content of this journal is prohibited without publishers
permission.
This
article is available in its entirety by fax for $40.00.
Visa, MasterCard & Amex accepted.
To