Published:Journal of Chromatographic Science,
ISSN 0021-9665 Volume 44,
Number 10, November/December 2006, pp.625-630
Determination of Volatile Organic Compounds by
Headspace Trap
F. Barani[1], N. Dell’Amico[1], L. Griffone[1],
M. Santoro[2], and C. Tarabella[1]
[1]Agenzia Regionale Protezione Ambiente Ligure, ARPAL, Dipartimento
della Spezia, Via Fontevivo 21, 19100 La Spezia, Italy and
[2]Perkin Elmer Italia, Via Tiepolo 24, 20052 Monza, Italy
A new analytical method for the determination of
halogenated and aromatic volatile organic compounds in groundwater,
mineral water, and drinking water at concentrations ranging between
1–10000 ng/L is developed. A new type of headspace sampler
that combines static headspace sampling with a trap is used, yielding
very low detection limits and good repeatability without carryover
effects. An unexpected transformation of 1,1,2,2-tetrachloroethane
into trichloroethene is observed and explained.