Published: Journal of Chromatographic Science, Volume 34, Number 10, October 1996, pp. 465–467.

Gas Chromatographic Analysis of Oxygen and Argon at Room Temperature
S.S. Raj, R.K. Sumangala, K.B. Lal, and P.K. Panicker

A simple and accurate method for the analysis of oxygen and argon at room temperature with a dual-column gas chromatographic system is described. The system consists of a 5A molecular sieve (MS 5A) column and a combination column (MS 5A and palladium alumina catalyst). The MS 5A column is used to separate the air sample into two main peaks at room temperature: an argon and oxygen combined peak and a nitrogen peak. The combination column is used to separate the argon and oxygen peaks; the oxygen is eliminated due to the catalytic reaction with hydrogen (hydrogen is used as the carrier gas), so only the argon peak is eluted. The concentrations of oxygen and argon can be calculated by comparing the output of both columns.

Reproduction of editorial content of this journal is prohibited without publisher’s permission.

This article is available in its entirety by fax for $4.00 per page.
Visa or MasterCard accepted.

To order electronically click here
or call: 847-647-2900 ext. 1323
or fax request to: 847-647-1155.

Please indicate JCS volume and issue along with page numbers.

 

Site Map: Home | Current Issue | Subscribe |
| Back Issues | About Us | Meetings | Advertising |
| Books for Sale | For the Author | Links |
| Supplier Info |Search |